The identified issue involves the breakage of the motor shaft combined with an error in the Single Fault Safety (SFS) system, designed to control descent speed. An insufficient oil level in the SFS safety drum could compromise its braking function, which users cannot visually check.
This dual failure increases the risk of patient falls or sudden drops, potentially causing injuries of varying severity. There are also risks for caregivers, such as mild to moderate musculoskeletal injuries if they attempt to prevent a patient's fall.
The distributing company, Baxter, has developed inspection and correction procedures for the affected equipment. Repairs can be carried out by authorized technical personnel from Baxter or qualified companies.
The AEMPS, in collaboration with the Swedish manufacturer Liko AB, has issued information notices for healthcare professionals, social health workers, and users/caregivers. It is recommended to stop using lifts that have exceeded their 10-year lifespan and to contact Baxter with any doubts or issues.
Until the inspection is completed, alternative mobile lifts should be used if available. If continuing to use an affected device, strict safety measures must be followed: seek additional caregiver assistance, keep the patient at the lowest possible height, clear the transfer environment, and continuously monitor the patient.